5011 Resistance / Temperature Calibrator
- Description
- Reviews (0)
Description
The 5011 is a versatile, high accuracy calibrator that can be used in a wide range of applications across various industries. Primarily used as a programmable resistance / RTD source, the 5011 provides precision resistance with a best accuracy of 0.01% and 1mΩ resolution (50Ω to 1kΩ).
Internal options for increased capabilities can be fitted as per customer requirement. These include DC voltage and thermocouple simulation, DC current, and frequency. The 5011 can be used to cover a wide workload as a laboratory calibrator or be incorporated into an automated test system. A rack mount kit option is also available.
Simple Operation and Clear Visual Display
Front panel operation allows the user to quickly set the function and output required. Using the jog / shuttle dial deviation the user can finely adjust the output value as a percentage (+/- 9.99%). All this information is shown on an easy to read LED display.
As standard the 5011 uses a SCPI command structure for programming. It also supports the Time Electronics 9811/19/20 command set making it an ideal replacement for these legacy models.
Flexible Options
A range of internally fitted options are available for the 5011, providing added functionality and features. The DCV / Thermocouple option provides a DC voltage source (+/- 20V) and simulates thermocouple types K, J, T, R, S, N, E and B. With the DC current option fitted the 5011 can source up to 220mA. This makes the instrument ideal for accurate process control calibration. Also available is a 0.1Hz to 10MHz frequency option (Period 100ns to 10s). The output is variable with a best resolution setting of 0.1Hz.
EasyCal Software – Manage, Automate and Optimise the Calibration Process
Connect the 5011 to a PC/Laptop (via RS-232, GPIB, or USB) installed with EasyCal software and automate the calibration process. This provides increased speed of calibration and consistency of results. Easily produce calibration certificates and reports to ISO 9001, ISO 17025, and other international quality standards.
Thông số kỹ thuật
Range | Accuracy | Resolution |
---|---|---|
1Ω to 20Ω | 0.01% +/- 5mΩ | 1Ω |
20Ω to 99.999Ω | 0.01% +/- 5mΩ | 1mΩ/5mΩ* |
100Ω to 999.999Ω | 0.01% +/- 5mΩ | 1mΩ |
1kΩ to 9.999kΩ | 0.02% +/- 20mΩ | 1Ω |
10kΩ to 99.999kΩ | 0.01% +/- 1Ω | 1Ω |
100kΩ to 999.99kΩ | 0.01% +/- 10Ω | 10Ω |
1MΩ to 9.9999MΩ | 0.02% +/- 100Ω | 100Ω |
10MΩ to 120MΩ | 0.1% +/- 1kΩ | 1kΩ |
* Output Setting Resolution below 50Ω is 5mΩ
Temperature Coefficient | Less than 50ppm per °C |
Power Rating | 0.1 Watt per resistor |
Maximum Voltage | 250V |
Resistance Switch Time | < 250μs |
Operation Time | < 300ms |
End Resistance Variation | < 2.5mΩ |
RTD Type | Alpha Coefficient | Range | Accuracy |
---|---|---|---|
PT100 | 0.003850 | -180 to -100°C | 0.1°C |
PT100 | 0.003850 | -100 to 850°C | 0.05°C |
It should be noted that the accuracy of the PRT simulation is determined by the accuracy of the PRT tables (BS EN 60751) published by the British Standards Institute. The 5011 uses precise digital interpretation of the tables to output resistance values that are within the accuracies specified in the table.
Type | Range °C | Accuracy °C |
---|---|---|
J | -210 to 150 / 150 to 1200 | 0.15 / 0.3 |
K | -270 to 190 / 190 to 1250 | 0.5 / 0.6 |
T | -200 to 150 / 150 to 400 | 0.4 / 0.5 |
R | -50 to 800 / 800 to 1750 | 0.8 / 2.0 |
S | -50 to 850 / 850 to 1750 | 0.9 / 2.0 |
B | 100 to 1200 / 1200 to 1800 | 1.0 / 2.0 |
N | -270 to 260 / 260 to 1300 | 0.5 / 0.4 |
E | -50 to 1000 | 0.3 |
Cold Junction Compensation +/- 0.5°C (applies to ambient changes of less than +/- 1°C at 23°C).
The accuracy of the thermocouple simulation is determined by the accuracy of the 5011’s DC voltage function and the accuracy of the standard thermocouple tables (BS EN 60584-1) published by the British Standards Institute. The 5011 uses precise digital interpretation of the tables to output voltage levels that are within the accuracies specified in the table.
Range | Accuracy | Resolution |
---|---|---|
20mV | 100ppm + 4μV | 100nV |
200mV | 30ppm + 6μV | 1μV |
2V | 25ppm + 20μV | 1μV |
20V | 25ppm + 150μV | 10μV |
* Specifications apply from 10 to 100% of range. 10% over range.
Range | Accuracy | Resolution |
---|---|---|
200μA | 150ppm + 15nA | 1nA |
2mA | 100ppm + 40nA | 10nA |
20mA | 80ppm + 200nA | 10nA |
200mA | 80ppm + 3μA | 100nA |
* Specifications apply from 10 to 100% of range. 10% over range.
Range | 0.1Hz to 10MHz / 100ns to 10s |
Accuracy | 20ppm |
Warm up | 30 minutes to full accuracy |
Settling Time | Less than 5 seconds |
Standard Interfaces | GPIB (IEEE-488), RS-232, USB |
Temp Performance | Operating: 0 to 40°C, Full Spec: 23°C +/- 5°C, Storage: -10°C to 50°C |
Operating Humidity | < 80% non-condensing |
Operating Altitude | 0 to 3km. Non operating: 3km to 12km |
Line Power | 100 – 230V AC 50/60Hz. Power Consumption 60W typical, 80W Max. |
Dimensions | W450 x D272 x H152mm (18 x 11 x 7”) |
Weight | 7kg (15.4lbs) |
Country of Origin | UK |
Supplied With | User manual, RS-232 cable, and USB adaptor/cable |