• Mô tả
  • Đánh giá (0)

Mô tả

The 7077 is a high performance process calibrator module that combines source and measurement functions for thermocouples, PT100s, μV, mV, and mA. As a multifunction module the 7077 combines accuracy with simple operation, making it ideal for process plant applications.

Thermocouple Measurement and Simulation
Measure and simulate the temperature and mV characteristics of J, K, T, R, S, B, N and E thermocouples.

Cold Junction Compensation
The unit can be operated with or without internal cold junction compensation.

PT100 Measurement and Simulation
Based on 0.3850 alpha probe standard. Range is -200°C to 700°C.

Measurement and Source (μV, mV, and mA)
Measurement ranges are 0 to +/-30mV and 0 to +/-60mA. Source ranges are 0 to +/-80mV and 0 to +80mA.

Temperature Units Selection
The display can be easily changed from °C to °F. The equivalent μV (thermocouples) and ohms (PT100) can also be shown.

24V Process Loop Drive Mode
A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA (or a lower level if required).

Inching (Incrementing/Decrementing)
The 7077 has a general-purpose inching function. This adjusts the output in fixed increments of temperature (thermocouples only) or voltage or current. The set-up menu gives a the user a choice of three levels of increment i.e. 0.1, 1 or 10 for °C/°F, or 1, 10, or 100 μV/μA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output. This is especially useful for calibrating thermostat controllers that have tight specification on hysteresis.

Memory Recall and Step/Auto-Step Functions
Up to 10 values can be stored in the unit’s non- volatile memory and they can be recalled at any time. The user can also manually step through them in sequence using the step key. Continuous stepping (auto-step) is also available at any user selectable rate between 1 and 10 seconds/step.

Thông số kỹ thuật

Thermocouple Type Temperature Range (°C) Accuracy (°C)
J -200 to 580 0.7
K -200 to -150
-150 to 750
2.5
0.5
T -200 to 0
0 to 400
1.5
0.4
R 50 to 400
400 to 1750
3.0
1.5
S -50 to 100
100 to 1750
3.0
1.5
B 110 to 1000
1000 to 1800
3.5
1.5
N -100 to 890 0.6
E -50 to 400 0.4
Resolution 0.1°C or 0.1°F
Thermocouple Type Temp Range °C Accuracy °C
J -210 to 150
150 to 1200
0.15
0.3
K -270 to 190
190 to 1250
0.5
0.4
T -200 to 150
150 to 400
0.4
0.5
R -50 to 800
800 to 1750
0.8
2.0
S -50 to 850
850 to 1750
0.9
2.0
B 100 to 1200
1200 to 1800
2.0
3.0
N -270 to 260
260 to 1300
0.5
1.0
E -50 to 1000 0.3
Resolution 0.1°C or 0.1°F

Note: An additional correction representing the equivalent 1μV should be allowed for stray thermal emf effects (Temperature measure & simulation).

14 set temperature points (°C) Accuracy
-100, -50, -20, 0, 20, 50, 100, 200, 300, 400, 500, 600, 700, 800 0.1% of resistance value (typically 0.5 °C)
Range Accuracy Resolution
200 to 700 °C, 2 wire. 0.2% of resistance value (typically 0.7 °C) 0.2°C (or 0.2°F)
Range Resolution Accuracy Input Resistance
0 to +/- 30mV 10μV 0.05% of full scale ±1 digit 100kΩ
Range Resolution Accuracy Output Resistance
0 to 8mV 0.5μV +/- 4μV 10Ω
8 to 80mV 5μV 0.02% of full scale 10Ω
Range Resolution Accuracy Input Resistance
0 to 60mA 20μA 0.05% of full scale ±1 digit 0.5Ω
Range Resolution Accuracy
0 to 8mA 0.5μA +/-10μA
8 to 80mA 5μA 0.02% of full scale
Inching Three levels of increment, 0.1, 1 or 10 for °C/°F, and 1, 10, or 100 μV/ μA for voltage/current. The lowest of these represents the highest setting resolution and provides the most precise control of the output.
24V Process Loop Drive A process loop can be driven at 24V and up to 60mA by selecting the ‘Milliamp Source’ mode and setting it at 60mA
(or a lower level if required)
Memory recall & step 10 memory locations for non-volatile storage of values. Manual & AutoStep, rate adjustable from 1 to 10 sec/step

Cold Junction Compensation Accuracy 0.2°C. Resolution 0.1°C.
Operating Temperature -10 to 40°C (15 to 105°F)
Connections Industry standard 4mm screw terminals
Module Width 295mm (primary or secondary console fitting)
Optional Extras Calibration certificates traceable NPL and UKAS

Đánh giá

Chưa có đánh giá nào.


Hãy là người đầu tiên nhận xét “7077 Process Calibrator Module”